Electron beam welding & radiation protection. Do you need a consent?
Published: Sept 28, 2025
Prelim
The regulatory aspects of this blog article is intended to apply to those who operate electron beam welders in the UK. The technical parts will generally apply to electron beam welders operated anywhere in the world.
Electron beam welding is a well known and established specialised technique of fusion welding which can be used to join two or more sections of metal together, or in some cases change the properties of a surface in some way. It is not the intention of this blog article to discuss the technique from a welding perspective, or describe an electron beam welder in significant detail. The essential parts of the welder that can be highlighted are as follows.
- A source of electrons - via a filament (electrons created by thermionic emission in a tungsten filament operating at a few hundred volts at most).
- A cathode (often part of the filament) which is at a negative potential compared to the anode (see below)
- An anode, and with the cathode creates the electric field which accelerates the electrons in the direction of the anode. The accelerating potential can vary between 30 - 200 kV (60 kV & 150 kV seem popular choices) and at 10's of mA (so power can easily be in the several kW range). Importantly unlike an x-ray tube (which so far sounds similar in description), the aim is not to intersect the anode, but rather pass through it (via an aperture within the anode - sometimes known as a donut anode).
- Sets of focusing coils / deflection coils which are used to focus the beam into a fine point onto the work object to be welded.
- A vacuum chamber which holds tooling and the object undergoing welding. For very large objects electron beam welding equipment can be configured such that it forms a vacuum seal against the object being welded. A vacuum (or partial vacuum) ensures that nothing interferes with the electron beam as it is formed.
- A viewing window (sometimes) - made of lead glass.
- Interlocks and other safety and process relate systems (more of the safety aspects coming up later).
The objective is to focus the electron beam onto the work object and form the welding process by converting the energy of the electrons (in keV) into thermal energy. Many W/mm2 (typically 104 - 107) of energy can be imparted over a small volume of material and the heat created used in the welding process.
The issue with respect to radiation protection arises from two main aspects:
- The creation of a high energy electron beam - this beam itself forming a source of ionising radiation (this overall is a minor significance - for example even high energy electrons are relatively easily stop in matter).
- The creation of adventitious x-rays by the process of bremsstrahlung - this being a by-product of the process (but may sometimes be utilised by an imager / detector to monitor weld quality).
The x-ray production is not as efficient as an x-ray tube (which itself is not an efficient process since most of the energy is expended in the form of heat), but a significant yield is still created which needs to be mitigated with shielding and other radiation protection techniques. Overall - when using an electron beam welder you are operating a radiation generator, even if that it not what you intend!
Do you need a consent?
Before looking at the potential radiation risks and exposure potential from operating an electron beam welder, and comparing with other radiation generators (e.g. an airport x-ray cabinet security system), we need to tackle the issue of registrations and consents - which are required in particular circumstances due to the Ionising Radiations Regulations 2017 (IRR17) in the UK.
We have already provided some guidance such as the following (will all open in a new tab).
- IRR17 (5) - Notification of certain work
- IRR17 (6) – Registration of certain practices
- IRR17 (7) – Consent to carry out specified practices
- X-ray cabinet sterilisation (irradiation) – UK registration or consent (IRR17)? [blog article]
- New UK Consent process for users of Ionising Radiation (as of February 2023) (Updated January 2025) [blog article]
Notwithstanding the above guidance and comment, we can summarise the requirements as follows:
- Notification of certain work - lowest level of radiological risk (e.g. for limited quantities of radioactive materials, occupational exposure to radon).
- Registration of a certain practice - medium level radiological risk (e.g. using a radiation generator such as an airport security x-ray machine).
- Consent to carry out specified practices - highest level of radiological risk (e.g. operation of an accelerator, industrial irradiation, use of high activity sealed sources etc).
The above is intended to represent the graded approach to radiation protection as outlined in IRR17, and also the likely degree of regulation required from HSE.
[Ionactive comment: There will hopefully be a 10 year review of IRR17 during 2027 (it is desperately needed) and we think the graded approach will need some significant improvement. For example, notification (lowest risk level) is required for occupational exposure to radon gas (where radon concentrations are > 300 Bq/m3). Yet radon gas probably yields the highest occupational exposures to any group of workers (or other persons) in the UK, whilst also being the most undeclared exposure of all - partly due to ignorance. Radon exposure can be serious - take a look at the following Ionactive blog article (will open in a new tab): Radon gas in schools (and other workplaces). The exposures derived in this blog article are way above those received by most workers who knowingly work with ionising radiation in the UK. ]
With the consent - what matters is that a particular specified practice is taking place, the type of ionising radiation source matters not. So industrial irradiation could use a radiation generator, an accelerator or radioactive sources - the relevant point is only that a specified practice is taking place.
The definition of industrial irradiation is found in IRR17 (2) and states specifically 'the use of ionising radiation to sterilise, process or alter the structure of products or materials'. It matters not what the equipment is, the physical aspects of the work, the shielding employed, the surface dose rates in normal use, whether a radiation accident is reasonably foreseeable, the output of a radiation risk assessment required by IRR17 (8). All that matters is that a specified practice is taking place.
It is true that some forms of industrial irradiation (such as large scale industrial sterilisation plants) carry significant radiation hazards and have massive radiation protection infrastructure. The relevant plants will have walk-in irradiation cells (e.g. for maintenance access) and require multiplicative and diverse safety systems, since they are capable of delivering say 50 kGy to items. But as already outlined, industrial irradiation cares not what the source of ionising radiation is - there is simply an assumption (?!) that in the worse reasonably foreseeable case, it could yield high unplanned radiation exposures.
So now we come to electron beam welding. A radiation generator (electron beam / secondary x-rays from bremsstrahlung) is used to alter the structure of a material (there can be little doubt about that). Therefore, under IRR17, electron beam welding is a form of industrial irradiation (a specified practiced) and requires a consent. The type of equipment, the kV, the mA, the surface dose rate, the output of the risk assessment is not relevant when making this declaration.
So - do you need a consent? If you do not have a consent you should not be operating an electron beam welder in the UK.
This is not just an Ionactive interpretation, the HSE state this clearly here: Work with ionising radiation that requires consent (will open in a new tab). Under examples for industrial irradiation is clearly states: 'This applies where you: work with a particle accelerator to perform ion implantation or electron beam welding'.
Now we will leave you to debate this with the HSE as their wording is not particularly helpful here. For example.
- Accelerator IS defined in IRR17 (2) as 'apparatus or installation in which particles are accelerated and which emits ionising radiation with an energy higher than 1 MeV'. However, particle accelerator (which the regulator uses above) is NOT defined in IRR17. Is there a difference?
Please don't get your hope up though - further up the paragraph of the above link, the following is stated:
You must apply for consent if you perform industrial irradiation using:
radioactive sources (for example HASS)
accelerators
radiation generators
There is no doubt that an electron beam welder is a radiation generator. IRR17 (2) defines a radiation generator as: 'a device capable of generating ionising radiation such as x-rays, neutrons, electrons or other charged particles'.
In case the above linked text is changed for whatever reason, the graphic below shows the website wording as of October 2025.
HSE website (October 2025) - work with ionising radiation requiring a consent
Given the above analysis and clear requirement for a consent, a brief review of the web revealed an alarming lack of obvious passive or active radiation warning signs during a picture search. It may be that such features are tucked away and simply not of interest to those taking pictures of electron beam welders! However in the UK we would expect to see passive x-ray warning signage and at least a "beam on" / "X-ray on" active signage visible when working around the unit. And perhaps a copy of the local rules.
Below are a few examples of what we might expect to see.
Electron beam welders - expectations in the UK?!
A basic risk comparison
We think it is fairly clear (from the above analysis) that operating an electron beam welder needs a consent. Before we consider what this means for the end user, we will spend a few moments comparing radiation risk between an electron beam welder and an airport security cabinet x-ray machine (for checked luggage). This does not change our interpretation of IRR17, but does lend weight to the debate that might take place if IRR17 is reviewed in 2027. We will make this comparison by way of a simple table. This is a qualitative example and not as rigorous as might otherwise be required if this was a formal assessment.
A radiation risk comparison (images are for illustration only)
Not all the attributes below are necessarily an advantage, and some would be prohibited.
| Attribute | Electron beam welder | Airport checked baggage x-ray machine (CT) |
|---|---|---|
| Enter interior for maintenance (radiation off) | Would be possible to enter vacuum chamber of larger units | Possible to enter interior of cabinet (e.g. for cleaning or clearing a bag jam) |
| Enter interior where radiation (x-rays) are being generated | Not generally possible - vacuum chamber sealed up | Entirely possible to enter unit from either end whilst x-rays are generated (prohibited) |
| Internal dose rate estimate | 20 Gy/h (see next section for estimate) | 1.8-3.6 Gy/h (based on 1-2 mSv per bag with 1800 bags per hour at 0.5 m/s conveyor speed). In beam instantaneous dose rate likely higher than this. |
| Likely exterior dose rate | < 1 μSv/h on surface (or at least at 10 cm from surface) | < 1 μSv/h on surface (or at least at 10 cm from surface). Instantaneous dose rates of 1-10 μSv/h possible at entrance & exit as shielding curtains move. |
| Ease of access whilst x-rays are generated | Not practicable (sealed vacuum chamber, high voltage, dangerous welding fumes, practical space etc) | Entirely possible. Each end of unit is open, protected by sets of flexible shielding curtain material (prohibited) |
| Use of interlocks to restrict access / inhibit x-rays generation | Yes. Interlocks used for multiple reasons - vacuum vessel needs to be proved shut, cannot be opened whilst vacuum present, high voltages, fumes, mechanical movements etc | Partially. Cover interlocks will inhibit user access to CT gantry and very high dose rate areas of unit. Interlocks will not impede access into the unit (baggage screening tunnel) when x-rays are being generated (but prohibited) |
| Controlled area exists inside unit when system is 'on' | Yes (HSE will insist this is the case). No reason to enter area with system energised (and not practically possible) | Yes (HSE will insist this is the case). No reason to enter area with system energised (& with x-rays on) - but entirely possible to do this (prohibited) |
| Local rules in place | Local rules (IRR17) will be required since a controlled area exists | Local rules (IRR17) will be required since a controlled area exists |
| Annual dose investigation level | Annual dose investigation level (DIL) needs to be stated in the local rules | Annual dose investigation level (DIL) needs to be stated in the local rules |
| Passive or active dosimetry | Not required for routine use of the equipment | Not required for routine use of the equipment |
| Periodic radiation survey around unit. | Required (not least to help prove DIL cannot be exceeded). Normally looking for < 1 μSv/h in areas that are accessible | Required (not least to help prove DIL cannot be exceeded). Normally looking for < 1 μSv/h in areas that are accessible. Note this will be time averaged, since higher IDR (see above) will be measurable at entry and exit |
| Radiation Protection Supervisor | Yes | Yes |
| Radiation Protection Adviser | Yes | Yes |
| Registration or Consent | Consent | Registration |
The above table is not meant to be exhaustive - but you get the idea. Many readers of this blog article would agree that if the designations of consent or registration were based purely on a risk assessment via IRR17 (8), then the last entries in the above table might well be swapped over. Considering routine use of either radiation generator, and accounting for reasonably foreseeable radiation incidents (or radiation accidents as defined in IRR17), which of the two units above do you think presents the higher radiological risk? A can of 🪱🪱!
[Ionactive comment: used compliantly, neither unit presents a significant radiation exposure risk.]
Ionising radiation generated when operating an electron beam welder
Before considering the implications of needing a consent, and in support of the above risk comparison, we will delve a little deeper into the ionising radiation generated within a typical electron beam welder. This is not a wasted exercise either, since a safety assessment (SA) required to apply for a consent, would certainly be enhanced using this type of analysis.
We are going to assume the following parameters/ settings:
- 150 kV acceleration (and also 60 kV).
- 40mA current (and also 60 mA at 60 kV).
- 6 kW power / 3. 6kW.
- Vacuum chamber (parts for welding are held in place within a vacuum chamber).
The above figures are for a pretty significant electron beam welding system, but the lower power systems (e.g. 60 kV at 60mA = 3.6 kW) will also be considered.
Being exposed to a 150 keV electron beam of this power would not be advisable (after all, it is being used to weld metal!). Such exposures are not reasonably foreseeable - for the vast majority of cases it takes place in a vacuum vessel. In addition, electron beams are poorly penetrating (i.e. easily shielded) so are not the dominant ionising radiation hazard in such equipment. We will therefore concentrate on the bremsstrahlung x-ray component, as this radiation will be far more penetrating and will need to be shielded.
We will borrow some resource from an Ionactive article (written for us by Dr Chris Robbins): Calculate an estimate of x-ray dose rate from an x-ray tube given kV and mA (opens in a new tab), but note we are not suggesting that the electron beam welder uses an x-ray tube (it does not).
The electron beam welder is not a particularly efficient producer of x-rays (it is not optimised for this purpose). The bremsstrahlung x-ray fraction of the total energy input can be approximated by as \[ F_e = 1 \times 10^{-6} Z E \] where:
- E is the energy in keV.
- Z is the atomic number.
The table below shows the atomic number (Z) for a few materials that may undergo electron beam welding, as compared to tungsten (W) which is the most common anode material used in industrial x-ray machines. The value Fe (bremsstrahlung x-rays) is also given. The table clearly shows that whilst x-ray generation from an electron beam (for the energies considered) is not that efficient for an x-ray tube, it is much less efficient for electron beam welding.
Element | Fe | Z number | E (keV) | % Brem |
|---|---|---|---|---|
Mg | 0.0018 | 12 | 150 | 0.18 |
Al | 0.00195 | 13 | 150 | 0.195 |
Ti | 0.0033 | 22 | 150 | 0.33 |
Fe (steel) | 0.0039 | 26 | 150 | 0.39 |
W | 0.0111 | 74 | 150 | 1.11 |
The above table does not take into account characteristic x-rays which are less important for this analysis.
It can be shown (gross approximation), that the dose rate in Gy/h can be estimated if we know the current, voltage and target material (being welded - not the anode) using the following expression. \[ D_{Gy hr^{-1}} = \frac{9 \times 10^{-7} \times I V^2 Z}{\pi r^2} \left( { \frac {\mu}{\rho}} \right) \]
The mass energy attenuation coefficient \( ({\mu}/{\rho}) \) is in units of \( m^2 kg^{-1}\), I is in amps, V is in volts and r is in meters (from the welding object). The expression is derived in the article linked above (the derivation will not be considered further here). For simplicity, in this assessment we assume average energy occurs at approximately 1/3 peak energy and is 57.3 keV for 150 kVp. If we calculate the dose rate beyond the confines of the vacuum (ignoring the shielding provide by the chamber for the moment) then we can estimate a dose rate in air using an air absorption coefficient of \(3.29 \times 10^{-3} \space m^{2}kg^{-1} \).
The above expression is then used to construct the following table which provides an estimated dose rate in Gy/h at 1m for the different target materials. [Ionactive comment: each target material will significantly alter the x-ray spectrum which should be considered if this analysis was a formal assessment. However, all we are looking for is some reasonable estimation which provides an order of magnitude likely dose rate which we can then consider further by adding some shielding. This simple analysis is also used to demonstrate that the electron beam welder is a radiation generator - even though this is not the intended purpose.
Indicative absorbed dose rates (in air) are given below for each material (150 kV / 40mA).
Element | Z Number | Voltage (kV) | Current (mA) | Dose rate (Gy/h) at 1m |
|---|---|---|---|---|
Mg | 12 | 150 | 40 | 10.18 |
Al | 13 | 150 | 40 | 11.03 |
Ti | 22 | 150 | 40 | 18.66 |
Fe (steel) | 26 | 150 | 40 | 22.05 |
W | 74 | 150 | 40 | 62.76 |
For comparison, indicative dose rates are given for the same materials at 60kV and 60 mA as shown in the table below.
Element | Z Number | Voltage (kV) | Current (mA) | Dose rate (Gy/h) at 1m |
|---|---|---|---|---|
Mg | 12 | 60 | 60 | 9.17 |
Al | 13 | 60 | 60 | 9.93 |
Ti | 22 | 60 | 60 | 16.81 |
Fe (steel) | 26 | 60 | 60 | 19.87 |
W | 74 | 60 | 60 | 56.54 |
[Ionactive comment: It may surprise you that the dose rates in air at 60 kV and 150 kV appear similar?! This is true, but rather academic once shielding is applied as noted shortly below. At 60 kV the average energy of the bremsstrahlung x-rays will be about 30 keV and here the mass energy attenuation coefficient in air rises considerably - not surprising if you think about. The data above will shortly be combined with shielding and the headline dose rates will be mitigated swiftly, especially those from the 60 kV scenario. If we were trying to measure these dose rates (if that were even possible), we might want to look at tissue equivalent doses which we would modify to yield effective whole body dose rate. For the purposes of this article we are just comparing values, so absorbed dose in air is fine.]
Radiation shielding
For this analysis we will use the dose rates calculated above at the reference distance of 1 m. 1 m is likely too large for the smallest of low kV welding unit, and too small for the largest of high kV units. Therefore 1 m seems a reasonable compromise. We will first look at a 60 kV scenario and then move on to a 150 kV scenario.
The design aim is that surface dose rates on the outside of the electron beam welder will not exceed 1 μSv/h. For our purposes we will assume that 1 μGy/h is ≈ 1μSv/h. The calculated dose rate will still be in air, but could be expressed as effective dose if desired in a formal calculation.
To determine shielding we will use the following Ionactive calculator: Diagnostic X-ray Shielding / Transmission Calculator (opens in a new tab). The referenced link explains how the calculator is used. For each kV noted above:
- We will take the dose rate for Fe (steel) - i.e. highest likely dose rate from the materials considered.
- Set the maximum shielded dose rate at 1 μSv/h (1 μGy/h) in the calculator.
- Output the required shielding to achieve this using lead or steel.
Without pre-empting the results too much at this point, it is likely that steel shielding will suffice for 60 kV, whereas lead will probably be needed at 150 kV. [Ionactive note: We understand it is possible to electron beam weld tungsten but this appears to be a very specialised procedure using extremely high power systems. Therefore we have stuck with steel since it is far more common and the shielding will accommodate the other materials in the above tables. If you are still curious about tungsten then use the dose rates in the above tables and plug them into the shielding calculator.]
60 kV shielding
Using the Ionactive shielding resource and the above criteria we have :
- 11.4 mm steel.
- 1.9 mm lead.
150 kV shielding
Using the Ionactive shielding resource and the above criteria we have :
- 96 mm steel.
- 7.2 mm lead.
It is most likely that a 150 kV system will use lead shielding (in form of a steel / lead hybrid shield).
The images below show the output of the calculator for each of the results above.
Overall this section concludes that radiation shielding from x-rays is significant. This is partly due to the radiation flux created by a high current (e.g. 40-60 mA), as compared to airport security x-ray equipment or even industrial radiography where currents of > 10mA are rare (for similar kV) . Whereas the much lower Z number of typical welding materials are less efficient in producing bremsstrahlung x-rays, the higher current in electron beam welding will to some degree compensate for this by creating an overall higher flux.
The above data does not specifically demonstrate a need for a consent, it is not spectacular compared to many other radiation generators that only need a registration. Recall that it's the specified practice (industrial irradiation) that drives the category, and not the specific radiological risks in a particular case. The radiological risk and dose assessment has no influence on the choices made in the graded approach of IRR17.
A consent for operating an electron beam welder
In this section we will consider what is required to apply for a consent to operate an electron beam accelerator. As part of the application process the employer will need to prepare a safety assessment (SA). The SA will require information on the following (seek Radiation Protection Adviser advice - you will need to appointment one anyway).
[Ionactive comment - this is not formal advice.]
| SA Section | Ionactive Comment |
|---|---|
| (1) A summary of the type of industrial irradiation performed and where this takes place. Details of the radiation generator must be given. | Include a general description - make clear it is electron beam welder. Provide the maximum kV and mA. If known, provide the maximum dose rate that can be delivered (within the unit). Provide maximum dose rate detectable on the surface of the enclosure / vacuum vessel. |
| (2) A summary of the arrangements for managing radiation protection during industrial irradiation. | Make clear what your company does. The assumption from the regulator might be that you are a large and complex user of ionising radiation (since a consent is being applied for). Detail the organisation (perhaps use a diagram). |
(3) Details of the likely frequency of use of the industrial irradiation sources of radiation to be used, or likely to be used. Projected maximum number of occasions industrial irradiation will be carried out per year at each location. Include accumulation of radon in the working environment and any other sources of radiation. | Frequency of use is likely related to the larger irradiators (using high activity sealed sources, or accelerators). Look at your workload and provide examples of the number of welds conducted in a day / week / month etc. Inclusion of the radon question occurs in all consent (and registration) applications. Whilst nothing to to with your intend use of an electron beam welder (radiation generator), it is recognised as a significant contributor to occupational exposure in some parts of the UK - hence HSE will take the opportunity to make you think about radon. |
| (4) An estimate of the radiation dose rates to which anyone can be exposed (employees, members of the public etc). Confirm dose rates will be < 7.5 micro Sv/h outside the shielding of enclosures. Consider where the exposure may occur (eyes, extremities, whole body). Provide evidence that all exposures are ALARP. | Recall that the consent application is a 'one size fits all' for industrial irradiation. For the electron beam welder, the short answer is - there will be no exposures to any class of person, measurable above background during routine use, and following any reasonably foreseeable unplanned event. You may wish to state that exposures will always be < 1 micro Sv/h at 10 cm from the surface (read the product literature). You may want a Radiation Protection Adviser (RPA) to conduct a formal radiation survey around your unit. |
| (5) A summary of the engineering control measures and design features already in place or planned (include all these features on a plan). | The wording and inference for this part of SA application leans towards a complex industrial irradiator. You have a desktop or floor mounted electron beam welder unit - read the brochure, and talk to the supplier (and don't overcomplicate for the sake of it). Things to include could be the following:
If the supplier does not have suitable product literature, you may wish to provide an annotated diagram of the electron beam welder, with the above features highlighted. It is probably worth denoting the interior of the system as a controlled area (HSE will expect this). |
| (6) A summary of the maintenance and testing schedules (for all features specified in section 5). | The supplier of the electron beam welder should provide information on the routine maintenance required (e.g. annual maintenance visit plus operator daily checks if relevant). Use this information to write an answer for this section. |
| (7) An summary of the results of any critical examination, or planned critical examination including pass / failure criteria). | One assumption is that you will not already have an electron beam welder (unless you have a consent already under the former system pre-RADAN, and have been asked to upgrade). Another assumption is that you have the system and do not already have a consent (for some reason ...). You will not be able to operate the electron beam welder until you have a consent. If you have been contacted by HSE out of the blue and it is agreed you do not have a consent, the regulator may allow you to continue to operate whilst you apply for a consent - this is not guaranteed. It is possible it's been installed prior to obtaining a consent, and therefore has a valid critical examination (by the supplier who should have their own consent). Either include the full critical examination report (if available), or ask the supplier to detail what the critical examination involves (they may provide a blank pro forma report detailing the checks and pass / failure criteria). |
| (8) A summary of the planned radiation dose rate monitoring regime for irradiation areas and their surroundings, including any areas to which the public may have access. | Whilst not all users of an electron beam welder will conduct their own routine radiation monitoring (i.e. leaving it to the RPA / service engineer etc), since this is an industrial irradiation consent, there will be an HSE expectation that the operator will do their own routine monitoring. So be prepared to purchase a suitable radiation dose rate monitor - the actual monitoring process will be simple, verifying, for example, that < 1 micro Sv/h at 10cm from the surface remains the minimum standard to achieve. You will need to detail the monitoring schedule, how records will be maintained, and the actions to be taken if the dose rate criteria is exceeded (switching off and unplugging is a good start!). |
| (9) A summary of the planned personal dosimetry to be provided to employees and others. | There will be no personal dosimetry required for operation of the electron beam welder. This should be clearly demonstrated in the IRR17 risk assessment. Note that routine personal dosimetry is not required for nearly all x-ray cabinet systems, including those with open ports (e.g. x-ray security screening units). Since a controlled area will be designated in the interior of the vacuum vessel, and therefore local rules will be required, it follows that a dose investigation level (DIL) will be required. Since personal dosimetry is not likely to be required, it follows that a dose rate monitoring regime (8) is a likely. |
| (10) The rationale for designating employees as classified persons. | There will be no requirement for any operator of the electron beam welder to be designated as a classified person. This should be clearly demonstrated in the IRR17 risk assessment. |
| (11) A summary of the radiological protection training that will be, or has been provided, to employees and other persons, including planned frequency and refresher training. | Radiation protection related training will be required as follows:
Such training should be refreshed over a period of 3-5 years. |
| (12) A summary of the information that will be supplied to employees concerning their work with ionising radiations in connection with pregnancy and breast feeding and how you will communicate this information to them. | Ionactive always advises that reassurance information / training (some of which will be presented in the local rules), is always available with respect to pregnancy. A radiation risk assessment for an electron beam welder, will show that there would be no restriction of work by an operator of the unit who declares they are pregnant. Information (and restrictions) regarding breast feeding is irrelevant with regards use of the an electron beam welder (no radioactive material is involved) - they may be non radiation reasons for restrictions. Whilst the above advice is obvious, still show this is true by considering this specifically in a radiation risk assessment. |
| (13) A summary of possible radiation accident situations as identified in the radiation risk assessment, their likelihood and potential severity. | [Ionactive note - the definition of radiation accident is as follows '...where immediate action would be required to prevent or reduce the exposure to ionising radiation of employees or any other persons...'] In our view there are no reasonably foreseeable radiation accidents - and therefore a contingency plan is not likely to be required. This should be clearly justified in the radiation risk assessment. Disconnection from the power supply will terminate x-rays whatever the circumstances. There may be reasonably foreseeable radiation related events (following evaluation by risk assessment). Examples could include:
If any of the above are deemed reasonably foreseeable they should be mitigated via contingency arrangements (don't call it a plan as that infers you have a radiation accident). |
The above is just the safety assessment (SA). You will also need the following in place:
- A programme of monitoring / auditing the arrangements you have in place for compliance with the consent and IRR17.
- A person with appropriate authority needs to be identified.
- A radiation risk assessment needs to be prepared (at least following para 70 and 71 of the ACoP to IRR17).
- A description of the engineering controls, design features, and safety related features to restrict radiation exposures to ALARP levels.
- One or more appointed Radiation Protection Supervisors (RPS) - suitably trained.
- A Suitable Radiation Protection Adviser (RPA) appointed in writing.
- Local rules have been prepared
- etc
If a UK electron beam welder user has all the above in place then - good! If you are using an electron beam welder with some or all of the above missing, then you have some considerable work to do. Seek advice from a Radiation Protection Adviser (RPA).
Mark Ramsay
Radiation Protection Adviser
Ionactive Consulting LTD
October 2025